Automated Versus Manual Blood Pressure Measurement: A Randomized Crossover Trial in the Emergency Department of a Tertiary Care Hospital in Karachi, Pakistan: Are Third World Countries Ready for the Change?

Authors

  • Kanaan Mansoor
  • Saba Shahnawaz Dialysis Unit, The Kidney Centre, Karachi
  • Mariam Rasool Sindh Government Hospital, Karachi
  • Huwad Chaudhry New Mehran Medical Centre, Karachi
  • Gul Ahuja Dr. Ziauddin University Hospital, Karachi
  • Sara Shahnawaz Dr. Ziauddin University Hospital, Karachi

DOI:

https://doi.org/10.3889/oamjms.2016.076

Keywords:

Sphygmomanometers, Automated oscillometer, blood pressure

Abstract

BACKGROUND: Hypertension has proven to be a strong liability with 13.5% of all mortality worldwide being attributed to elevated blood pressures in 2001. An accurate blood pressure measurement lies at the crux of an appropriate diagnosis. Despite the mercury sphygmomanometer being the gold standard, the ongoing deliberation as to whether mercury sphygmomanometers should be replaced with the automated oscillometric devices stems from the risk mercury poses to the environment.

AIM: This study was performed to check the validity of automated oscillometric blood pressure measurements as compared to the manual blood pressure measurements in Karachi, Pakistan.

MATERIAL AND METHODS: Blood pressure was recorded in 200 individuals aged 15 and above using both, an automated oscillometric blood pressure device (Dinamap Procare 100) and a manual mercury sphygmomanometer concomitantly. Two nurses were assigned to each patient and the device, arm for taking the reading and nurses were randomly determined. SPSS version 20 was used for analysis. Mean and standard deviation of the systolic and diastolic measurements from each modality were compared to each other and P values of 0.05 or less were considered to be significant. Validation criteria of British Hypertension Society (BHS) and the US Association for the Advancement of Medical Instrumentation (AAMI) were used.

RESULTS: Two hundred patients were included. The mean of the difference of systolic was 8.54 ± 9.38 while the mean of the difference of diastolic was 4.21 ± 7.88. Patients were further divided into three groups of different systolic blood pressure <= 120, > 120 to = 150 and > 150, their means were 6.27 ± 8.39 (p-value 0.175), 8.91 ± 8.96 (p-value 0.004) and 10.98 ± 10.49 (p-value 0.001) respectively. In our study 89 patients were previously diagnosed with hypertension; their difference of mean systolic was 9.43 ± 9.89 (p-value 0.000) and difference of mean diastolic was 4.26 ± 7.35 (p-value 0.000).

CONCLUSIONS: Systolic readings from a previously validated device are not reliable when used in the ER and they show a higher degree of incongruency and inaccuracy when they are used outside validation settings. Also, readings from the right arm tend to be more precise.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

Plum Analytics Artifact Widget Block

Author Biography

Kanaan Mansoor

Kanaan Mansoor, MBBS, MD, CMO,Emergency Deparment , Dr. Ziauddin University Hospital, Karachi, Pakistan.

References

Kaczorowski J, Dawes M, Gelfer M. Measurement of blood pressure: new developments and challenges. Blood pressure. 2007:7.

Abboud H, Henrich WL. Stage IV chronic kidney disease. N Engl J Med. 2010;362:56-65. http://dx.doi.org/10.1056/NEJMcp0906797 PMid:20054047

Staessen JA, Richart T, Birkenhäger WH. Less atherosclerosis and lower blood pressure for a meaningful life perspective with more brain. Hypertension. 2007; 49:389-400. http://dx.doi.org/10.1161/01.HYP.0000258151.00728.d8 PMid:17283254

Faruqui A. Hypertension in Pakistan: What are we going to do about it? Pakistan Heart Journal. 1997;30:3-4.

Wilkins K, Campbell NRC, Joffres MR, et al. Blood pressure in Canadian adults. Health Rep. 2010;21:37-46. PMid:20426225

Lawes CM, Vander Hoorn S, Rodgers A. International Society of Hypertension. Global burden of blood-pressure-related disease, 2001. Lancet. 2008;371(9623):1513-1518. http://dx.doi.org/10.1016/S0140-6736(08)60655-8

Reeves RA. Does this patient have hypertension? How to measure blood pressure. JAMA. 1995;273:1211–1216. http://dx.doi.org/10.1001/jama.1995.03520390071036 PMid:7707630

Ogedegbe G, Pickering T. Principles and techniques of blood pressure measurement. Cardiology Clinics. 2010;28(4):571-586. http://dx.doi.org/10.1016/j.ccl.2010.07.006 PMid:20937442 PMCid:PMC3639494

O'Brien E. Has conventional sphygmomanometry ended with the banning of mercury? Blood Press Monit. 2002;7:37–40. http://dx.doi.org/10.1097/00126097-200202000-00007 PMid:12040241

O'Brien E, Asmar R, Beilin L, Imai Y, Mallion J-M, Mancia G, Mengden T, Myers M, Padï¬eld P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. Journal of Hypertension. 2003;21:821–848. http://dx.doi.org/10.1097/00004872-200305000-00001 PMid:12714851

Le Pailleur C, Helft G, Landais P, et al. The effects of talking, reading, and silence on the ''white coat'' phenomenon in hypertensive patients. Am J Hypertens. 1998;11:203–207. http://dx.doi.org/10.1016/S0895-7061(97)00403-2

Mancia G, Bertinieri G, Grassi G, Parati G, Pomidossi G, Ferrari et al. Effects of blood pressure measurement by the doctor blood pressure and heart rate. Lancet. 1983; ii:695–698. http://dx.doi.org/10.1016/S0140-6736(83)92244-4

Mancia G, Ferrari A, Gregorini L, Parati G, Pomidossi G, Bertet al. Blood pressure and heart rate variabilities in normotensive hypertensive human beings. Circ Res. 1983; 53:96–104. http://dx.doi.org/10.1161/01.RES.53.1.96 PMid:6861300

Parati G, Ulian L, Santucciu C, Omboni S, Mancia G. Blood pressure variability, cardiovascular risk and antihypertensive treatment. J Hypertens Suppl. 1995;13(4): S27–S34. http://dx.doi.org/10.1097/00004872-199512002-00005 PMid:8824684

Parati G, Ulian L, Santucciu C, Omboni S, Mancia G. Difference between clinic and daytime blood pressure is not a measure white coat effect. Hypertension. 1998;31(5):1185–1189. http://dx.doi.org/10.1161/01.HYP.31.5.1185 PMid:9576133

Myers MG, Valdivieso M, Kiss A. Use of automated office blood pressure measurement to reduce the white coat response. J Hypertens. 2009;27:280–286. http://dx.doi.org/10.1097/HJH.0b013e32831b9e6b PMid:19155785

Myers, M. G. Why Automated Office Blood Pressure Should Now Replace the Mercury Sphygmomanometer. The Journal of Clinical Hypertension. 2010;12: 478–480. http://dx.doi.org/10.1111/j.1751-7176.2010.00301.x

Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Curry F, Grant, Kaczorowski J. Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial. BMJ. 2011; 342:d286. http://dx.doi.org/10.1136/bmj.d286 PMid:21300709 PMCid:PMC3034423

Skirton H, Chamberlain W, Lawson C, Ryan H, Young E. A systematic review of variability and reliability of manual and automated blood pressure readings. Journal of Clinical Nursing. 2011; 20: 602–614. http://dx.doi.org/10.1111/j.1365-2702.2010.03528.x PMid:21320189

World Health Organization. Department of Protection of the Human Environment Water, Sanitation and Health. Mercury in Health Care. August 2005.

Hospitals for a Healthy Environment. Regulated Medical Waste Reduction. http://www.h2e-online.org.

Health Care Without Harm. End of An Era. The phase-out of mercury-based blood pressure measurement devices in the United States and its implications for Europe and the rest of the world. 2008. http://noharm.org/lib/downloads/mercury/End_of_an_Era_Mercury.pdf

Myers MG, McInnis NH, Fodor GJ, Leenen FHH. Comparison between and Automated and manual sphygmomanometer in a population survey. Am J Hypertens. 2008;21(3):280-283. http://dx.doi.org/10.1038/ajh.2007.54 PMid:18219304

O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman D, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993;11(Suppl 2):S43–S63.

Heinemann M, Sellick K, Rickard C, Reynolds P, McGrail M. Automated versus Manual blood pressure measurement: A randomized crossover trial. Int J Nurs Pract. 2008;14:296-302. http://dx.doi.org/10.1111/j.1440-172X.2008.00696.x PMid:18715391

O'Brien E, Waeber B, Parati G, Staessen J, Myers MG. Blood pressure measuring devices: recommendations of the European Society of Hypertension. Brit Med J. 2001;322:531-536. http://dx.doi.org/10.1136/bmj.322.7285.531 PMid:11230071 PMCid:PMC1119736

Pickering TG, Hall JE, Appel LJ, Falkner B, Graves J, Hill M, et al. Recommendations for blood pressure measurement in humans and experimental animals, part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension. 2005;45:142-161. http://dx.doi.org/10.1161/01.HYP.0000150859.47929.8e PMid:15611362

British Hypertension Society http://www.bhsoc.org/bp-monitors/bp-monitors/

Hutton P, Dye J, Prys-Roberts C. An assessment of the Dinamap 845. Anaesthesia. 1984;39:261–267. http://dx.doi.org/10.1111/j.1365-2044.1984.tb07240.x PMid:6703297

Rose KM, Arnett DK, Ellison RC, Heiss G. Skip patterns in DINAMAP measured blood pressure in 3 epidemiological studies. Hypertension. 2000;35:1032–1036. http://dx.doi.org/10.1161/01.HYP.35.5.1032 PMid:10818059

Dind A, Short A, Ekholm J, Holdgate A. The inaccuracy of automatic devices taking postural measurements in the emergency department. International Journal of Nursing Practice. 2011; 17: 525–533. http://dx.doi.org/10.1111/j.1440-172X.2011.01958.x PMid:21939485

van Popele NM, Bos WJ, de Beer NA, van Der Kuip DA, Hofman A, Grobbee DE, Witteman JC. Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer. Hypertension. 2000;36(4):484-8. http://dx.doi.org/10.1161/01.HYP.36.4.484 PMid:11040223

Braam RL, Thien T. Is the accuracy of blood pressure measuring devices underestimated at increasing blood pressure levels? Blood Press Monit. 2005;10(5):283-9. http://dx.doi.org/10.1097/01.mbp.0000180671.76279.c7

Association for the Advancement of Medical Instrumentation. American National Standard. Manual, electronic or automated sphygmomanometers ANSI/AAMI SP10-2002/A1. Arlington, VA 22201-4598, USA: AAMI, 2003.

O'Brien E, Atkins N. A comparison of the British Hypertension Society and Association for the Advancement of Medical Instrumentation protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens. 1994; 12(9):1089-94. http://dx.doi.org/10.1097/00004872-199409000-00015

Wan Y, Heneghan C, Stevens R, et al. Determining which automatic digital blood pressure device performs adequately: A systematic review. Journal of Human Hypertension. 2010; 24: 431–438. http://dx.doi.org/10.1038/jhh.2010.37 PMid:20376077 PMCid:PMC2897978

World Bank – Poverty Data; http://povertydata.worldbank.org/poverty/country/PAK

Published

2016-07-27

How to Cite

1.
Mansoor K, Shahnawaz S, Rasool M, Chaudhry H, Ahuja G, Shahnawaz S. Automated Versus Manual Blood Pressure Measurement: A Randomized Crossover Trial in the Emergency Department of a Tertiary Care Hospital in Karachi, Pakistan: Are Third World Countries Ready for the Change?. Open Access Maced J Med Sci [Internet]. 2016 Jul. 27 [cited 2024 Apr. 23];4(3):404-9. Available from: https://oamjms.eu/index.php/mjms/article/view/oamjms.2016.076

Issue

Section

B - Clinical Sciences