Arthroscopic Suture Anchor Design Finite Element Study

Authors

  • Mai Ayoub Department of Mechanical Engineering, National Research Centre, Cairo, Egypt https://orcid.org/0000-0002-9360-9290
  • Mohamed EL-Anwar Department of Mechanical Engineering, National Research Centre, Cairo, Egypt
  • Mazen I. Negm Department of Mechanical Engineering, Faculty of Engineering, Al-Azhar University, Cairo, Egypt

DOI:

https://doi.org/10.3889/oamjms.2021.6409

Keywords:

Finite element analysis, Design, Arthroscopic anchors, Suture eyelet

Abstract

AIM: This in-vitro study investigated arthroscopic suture anchors’ main design parameters effect on surrounding bone.

METHODS: Thirty-dimensional arthroscopic suture anchor designs’ models were created on engineering CAD software by changing thread profile, pitch, and anchor tip profile as design parameters. These models were imported into ANSYS Workbench for finite element analysis. Bone was simplified and modeled as two coaxial cylinders. Tensile vertical load of 300 N, and oblique at 45º to the vertical axis, were applied to each model as two loading conditions while the simplified bone base was fixed in place as a boundary condition.

RESULTS: The finite element analyses on all models under both loading conditions showed stresses within physiological limits on bone. Trapezoidal teeth and inclined cut teeth designs showed the lowest values of stresses and deformations respectively on the bone under oblique loads, while curved tooth and square tooth designs showed the lowest values of stresses and deformations respectively on the bone under vertical loads. General ascending or descending trend was recorded by increasing pitch from 1.2 to 1.5 to 1.8 mm on the total deformation and maximum Von Mises stress on bone and anchor body. Tapered tip slightly increased bone and anchor stresses.

CONCLUSION: Arthroscopic anchors thread profile has minor affect on cortical bone behavior. Trapezoidal teeth, square tooth, and inclined cut teeth profiles showed the lowest values of stresses and deformations on cortical bone. Increasing thread pitch of arthroscopic suture anchors increases or decreases stress on the bone, and anchor body according to thread profile edges. Anchor tip profile negligibly affects both deformations and stresses on bone and anchor body.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

Plum Analytics Artifact Widget Block

References

Azato FN, Yamasaki AT, Sucomine F. Traction endurance biomechanical study of metallic suture anchors at different insertion angles. Acta Ortop Bras. 2003;11(1):25-31. https://doi.org/10.1590/s1413-78522003000100004 DOI: https://doi.org/10.1590/S1413-78522003000100004

Schwitalla A, Muller WD. PEEK dental implants: A review of the literature. J Oral Implantol. 2013;41(6):743-9. PMid:21905892 DOI: https://doi.org/10.1563/AAID-JOI-D-11-00002

Najeeb S, Khurshid Z, Matinlinna JP, Siddiqui F, Nassani MZ, Baroudi K. Nanomodified PEEK dental implants: Bioactive composites and surface modification-a review. Int J Dent. 2015;2015:381759. https://doi.org/10.1155/2015/381759 PMid:26495000 DOI: https://doi.org/10.1155/2015/381759

Suture Anchor Assembly, United States Patent No. 4, 632, 100; 1986.

Aktay SA, Kowaleski MP. Analysis of suture anchor eyelet position on suture failure load. Vet Surg. 2011;40(4):418-22. https://doi.org/10.1111/j.1532-950x.2011.00834.x PMid:21539579 DOI: https://doi.org/10.1111/j.1532-950X.2011.00834.x

Hughes CM. A Finite Element Modelling Strategy for Suture Anchor Devices. PhD Thesis, School of Engineering and Design, Brunel University; 2014.

Suture Anchor and Driver Assembly, United States Patent No. 5, 100, 417; 1992.

Bone Screw, United States Patent No. 5, 169, 400; 1992.

Suture Anchor Assembly, United States Patent No. 5, 370, 662; 1994.

Bone Screw with Improved Threads, United States Patent No. 5, 417, 533; 1995.

El-Anwar M. Second Annual report on Redesign of Some Endoscopic Instruments and Implants (GIT and Joints). Egypt: National Research Centre.

International Organization for Standardization. Safety Aspects- Guidelines for their Inclusion in Standards, ISO/IEC Guide No. 51; 2014.

International Organization for Standardization. Biological Evaluation of Medical Devices-Part 12: Sample Preparation and Reference Materials, ISO No. 10993-12; 2012.

International Organization for Standardization. Medical Devices-Application of Risk Management to Medical Devices, ISO No. 14971; 2007.

European Committee for Standardization (CEN). Medical Devices-Application of Risk Management to Medical Devices, EN ISO No. 14971; 2012.

International Organization for Standardization. Biological Evaluation of Medical Devices-Guidance on the Conduct of Biological Evaluation within a Risk Management Process, ISO TR No. 15499; 2012. https://doi.org/10.3403/30211401 DOI: https://doi.org/10.3403/30211401

Meyer DC, Nyffeler RW, Fucentese SF, Gerber C. Failure of suture material at suture anchor eyelets. Arthroscopy. 2002;18(9):1013-9. https://doi.org/10.1053/jars.2002.36115 PMid:12426545 DOI: https://doi.org/10.1053/jars.2002.36115

Barber FA, Herbert MA, Beavis RC, Barrera Oro F. Suture anchor materials, eyelets, and designs: Update 2008. Arthroscopy. 2008;24(8):859-67. https://doi.org/10.1016/j.arthro.2008.03.006 PMid:18657733 DOI: https://doi.org/10.1016/j.arthro.2008.03.006

El-Anwar M, Osman W. Finite element study on arthroscopic anchor design aspects. Open Access Maced J Med Sci. 2019;7(4):628-31. https://doi.org/10.3889/oamjms.2019.164 PMid:30894926 DOI: https://doi.org/10.3889/oamjms.2019.164

Wright PB, Budoff JE, Yeh ML, Kelm ZS, Luo ZP. Strength of damaged suture: An in vitro study. Arthroscopy. 2006;22(12):1270-5. https://doi.org/10.1016/j.arthro.2006.08.019 PMid:17157724 DOI: https://doi.org/10.1016/j.arthro.2006.08.019

Barber FA, Herbert MA, Click JN. Internal fixation strength of suture anchors-update 1997. Arthroscopy. 1997;13(3):355-62. https://doi.org/10.1016/s0749-8063(97)90034-7 PMid:9195034 DOI: https://doi.org/10.1016/S0749-8063(97)90034-7

Yakacki CM, Griffis J, Poukalova M, Gall K. Bearing area: A new indication for suture anchor pullout strength? J Orthop Res. 2009;27(8):1048-54. https://doi.org/10.1002/jor.20856 PMid:19226593 DOI: https://doi.org/10.1002/jor.20856

Schneeberger AG, Von Roll A, Kalberer F, Jacob HA, Gerber C. Mechanical strength of arthroscopic rotator cuff repair techniques: An in vitro study. J Bone Joint Surg. 2002;84A(12): 2152-60. DOI: https://doi.org/10.2106/00004623-200212000-00005

Downloads

Published

2021-07-22

How to Cite

1.
Ayoub M, EL-Anwar M, Negm MI. Arthroscopic Suture Anchor Design Finite Element Study. Open Access Maced J Med Sci [Internet]. 2021 Jul. 22 [cited 2024 Mar. 28];9(A):562-6. Available from: https://oamjms.eu/index.php/mjms/article/view/6409

Issue

Section

Sports Medicine

Categories

Funding data

  • National Research Centre
    Grant numbers This research was carried out via internal project entitled "Redesign of some Endoscopic Instruments and Implants (GIT and joints)" - code: 11090336